Table of Contents
1 Specifications
on page 3
2 Online user manual
on page 4
3 General information
on page 4
4 Installation
on page 5
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Ambient temperature
Storage temperature
Operating humidity
Operating altitude
Environmental
conditions
EMC requirements
Korean registration
Compliance
Pollution degree
Enclosure ratings
Power supply
Analog current output
version on the
measurement board
Analog voltage output
version on the
measurement board
Details
–5 to 50°C (23 to 122°F)
–20 to 70°C (–4 to 158°F)
0 to 95% non-condensing relative humidity
2000 m (6561 ft) maximum
Indoor use
EN61326-1: EMC Directive
Note: The wall mount instrument is a Class A product. In a domestic environment this product may
cause radio interference in which case the user may be required to take adequate measures.
User Guidance for EMC Class A Equipment
업무용을 위한 EMC 등급 A 장치에 대한
사용자 지침
사용자안내문
A 급 기기 ( 업무용 방송통신기자재 )
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기
바라며 , 가정외의 지역에서 사용하는 것을 목적으로 합니다.
CE, ETL certified to UL and CSA safety standards (with all sensor types), FCC, KC,
RCM, EAC, UKCA, SABS
2
IP65; Totally protected against dust; Protected against low pressure jets of water from all
directions.
NEMA 4X (wall mount only); Totally protected against dust; Protected against pressure
jets of water from all directions.
Enclosure rating does not apply to external power supply for benchtop instruments.
Universal 100 VAC to 240 VAC ±10% @ 50/60Hz - 40VA; 10 to 30 VDC - 30W
4-20 mA (default) or 0-20 mA (configuration with software);
3 configurable outputs; Maximum load: 500 ohm; Sensitivity: 20μA;
Accuracy: ± 0.5% (between operating temperature limits)
0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)