Descargar Imprimir esta página

Hach ORBISPHERE 410 Manual Básico Del Usuario página 11

Ocultar thumbs Ver también para ORBISPHERE 410:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 103
Figure 7 Cable gland type 2 wiring
1 Gland nut
2 Gasket
3 Cones
4 Spring
5. Reattach and tighten the cable gland nut.
6. Attach the wires to the corresponding terminal block connections.
4.6 Connection to mains power supply
4.6.1 Power supply connection (low voltage instruments)
For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 8-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
Figure 8 BINDER connector
4.6.2 Power supply connection (high voltage instruments)
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Electrocution hazard. Always remove power to the instrument before making electrical connections.
5 Shield
6 Tape
7 Gland body
8 Instrument
Pin Connections:
1. Power 10-30 VDC
2. Ground
3. Ground
4. Ground
5. Not used
6. Power 10-30 VDC
7. Power 10-30 VDC
8. Earth
D A N G E R
D A N G E R
9 O-ring
10 Wires
Figure 9 Wiring side view
English 11

Publicidad

loading