˜ Calibration
NOTES:
The thickness of the calibration plate 
o
range of 1 .5 mm to 2 .2 mm .
The initial display is 1 .80 mm . Press the TEST button 
o
to increase the reading on the display 
UNIT button 
The reading should equal the thickness value of the
o
calibration plate . For example, if the thickness of the
calibration plate is 1 .76 mm, press the UNIT button to
decrease the reading on the display to 1 .76 mm .
The thickness of the calibration plate should be greater than
o
the layer thickness on coated panel .
1 . Before a measurement in NFe or Fe test mode is made,
ensure that in this mode, the product has been calibrated
by using the naked substrate 
Alternatively a naked substrate can be used, if it is identical
to the substrate of the object to be tested .
to decrease the reading on the display .
4 ]
[
must be in the
7 ]
[
. Press the
2 ]
[
of the object to be tested .
8 ]
[
[
3 ]
GB/IE
53