User performance results for glucose concentrations between
48 mg/dL and 553 mg/dL.
Within ± 15 mg/dL and Within ± 20 %
Precision
Precision studies were performed in a laboratory using the
GLUCOCARD Shine Connex Blood Glucose Monitoring System.
Within Run Precision
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Between Run Precision
Control avg.
Control avg.
Control avg.
60 www.arkrayusa.com
371/371 (100 %)
43 mg/dL
71 mg/dL
135 mg/dL
203 mg/dL
343 mg/dL
36 mg/dL
114 mg/dL
341 mg/dL
SD = 1.9 mg/dL
SD = 2.1 mg/dL
SD = 3.8 mg/dL
SD = 5.2 mg/dL
SD = 11 mg/dL
SD = 1.4 mg/dL
SD = 3.4 mg/dL
SD = 8.2 mg/dL
CV = 4.3 %
CV = 2.9 %
CV = 2.8 %
CV = 2.6 %
CV = 3.2 %
CV = 3.8 %
CV = 3.0 %
CV = 2.4 %