If the equipment is used in a manner not specified by this document and K0553 the protection
provided by the equipment may be impaired.
1.3 Safety
Examine all equipment to be used, paying particular attention to electrical leads and connectors
and pneumatic pipes and connectors. DO NOT use any equipment that is known to be or
suspected to be damaged or faulty.
1.4 Power Supply
Single phase
1.5 Power Supply Connection
Do not apply electrical power outside the values stated.
Make sure that the power supply is off before connecting the power cable.
The ADTS must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
RISK OF ELECTRIC SHOCK
if there is not a protective earth in place. The ADTS must, at all times, be
connected to the supply protective earth/ground.
The power supply cable and connector must be correctly rated for the power
supply. Only use power cables supplied by Druck specifically for the ADTS.
European Color
Green/Yellow
1.6 Fuses
The two High Breaking Capacity fuses (8) (see Figure 6), located in the holders and mounted on
the front panel, protect the unit. The fuses are connected in the live and neutral supply circuit and
are rated at: T5AH 250V.
1.7 Replacing a Fuse
1.
Disconnect the electrical power supply.
2.
Unscrew the fuse holder cap from the front panel and remove the fuse.
3.
Fit a new fuse of the correct size and rating into the fuse holder cap.
4.
Secure the fuse holder cap in the front panel.
2. ADTSTOUCH
The ADTSTOUCH is used to control all required functions. The ADTSTOUCH can be positioned
on the ADTS or used as a hand held mobile unit via a cable or Bluetooth® wireless technology.
The ADTSTOUCH will have power applied when:
•
Positioned (docked) (1) on a powered-on ADTS.
•
Using an umbilical cable connected to a powered-on ADTS.
110/230 Vac, 50/60Hz
115 Vac, 400Hz
US Color
Brown
Blue
200 VA Max - ADTS542F/552F
300 VA Max - ADTS553F/554F
There is a risk of electric shock/electrocution
Black
White
Green
Protective Earth (Ground)
English–ADTS542F/552F/553F/554F Safety and Installation Guide | 3
Function
Live
Neutral
Copyright 2014 Baker Hughes Company.