Note: The battery does not necessarily need to be new or re-charged. The ADTSTOUCH
will function safely and correctly with a flat battery when the umbilical cable or dock charging
supply is connected.
4.
Refit the battery compartment cover plate and tighten the two thumbscrews (5). The cover
must be fitted correctly to maintain protection afforded.
2.2 Battery Care and Maintenance
The battery unit contains a Lithium-ion battery which requires no user maintenance. If shipping
an ADTSTOUCH and/or battery unit please contact your local service center beforehand for
shipping requirements. Battery must be removed from the ADTSTOUCH during transportation.
Prolonged exposure to temperature extremes may significantly reduce battery lifetime. For
maximum lifetime, avoid prolonged periods where the battery is exposed to temperatures outside
the range -30°C to +45°C.
The recommended storage temperature range is 5°C to 21°C (41°F to 98.8°F).
The battery is removable. The manufacturer makes the following safety recommendations. The
user should not:
Short circuit the battery.
Immerse the battery in any liquid.
Disassemble or deform the battery.
Expose to, or dispose of the battery in a fire.
Subject the battery to excessive physical shock or vibration outside the specified limits for
the ADTS.
Use a battery that appears to have suffered abuse.
Charge with an unspecified charger or specified charger that has been modified. This can
cause breakdown of the battery or swelling and rupturing.
Use a battery in an appliance or purpose for which it was not intended.
The battery unit is sealed for life, so leakage of electrolyte is not expected. If leakage of electrolyte
from the battery is observed, stop using the battery immediately and avoid contact with the
electrolyte. If skin or clothes come into contact with the electrolyte, wash immediately with soap
and water. If electrolyte comes into contact with the eyes, wash the eyes thoroughly with water
and consult a doctor immediately.
The ADTS542F/552F/553F/554F, ADTSTOUCH and ADTSTOUCH-ER contain a Lithium-ion
coin cell. This coin cell is not user replaceable. Installation and replacement shall only be
performed by trained personnel.
Always dispose of batteries safely and in accordance with local instructions.
2.3 Battery Charging
The battery in the ADTSTOUCH can be charged using one of the following methods:
By placing the ADTSTOUCH onto a powered ADTS.
By connecting the ADTSTOUCH connector (4) (see Figure 3) to the ADTS using the
umbilical cable.
By using the AATOUCH-5 optional accessory desktop charger. The battery must be
removed from ADTSTOUCH when it is charged by the AATOUCH-5 desktop charger.