Descargar Imprimir esta página

Siemens ICS Schneider Messtechnik CLS200 Manual De Instrucciones página 9

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69
Wiring
WARNING: All field wiring must have insulation suitable for at least
250 V.
Notes:
Only qualified personnel are authorized to install and operate this equipment in
accordance with established safety practices and standards.
The Protective Earth Terminal indicated by
All wiring must be done by qualified personnel in accordance with all governing
regulations.
The equipment must be protected by a 15A fuse or circuit breaker in the building
installation.
A circuit breaker or switch in the building installation, marked as a disconnect switch,
shall be in close proximity to the equipment and within easy reach of the operator.
Use shielded cable, wire gauge 20 AWG to 14 AWG (0.5 mm
with a braided metallic shield for CE installations (or armoured cable where applicable).
Maximum working voltage between adjacent relay contacts is 250 V.
Relay contact terminals are for use with equipment which has no accessible live parts
and wiring which has insulation suitable for at least 250 V.
1.
Loosen the lid clip and remove the lid to access the connectors and electronics. (The
diagram on the next page can also be found on the underside of the lid, together with a
guide to switch function).
2.
Connect the wires to the terminals (polarity is not important).
3.
Ground the instrument according to local regulations.
4.
Tighten the gland to form a good seal.
5.
After adjusting the settings, replace the lid and secure the lid clip.
A5E32221251
Pointek CLS200/300 (Standard) – QUICK START MANUAL
must be connected to reliable ground.
2
2
to 2.0 mm
). Use a cable
Page EN-7

Publicidad

loading

Este manual también es adecuado para:

Ics schneider messtechnik cls300