Adjusting The Adjustable Dp-Unit - MIETHKE proGAV 2.0 Instrucciones De Manejo

Ocultar thumbs Ver también para proGAV 2.0:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

proGAV 2.0
proGAV Check-mate
The proGAV Check-mate is delivered sterile
and is intended to be re-sterilised. It is possi-
ble to change and to verify an applied pressure
setting on the valve directly. To verify the actual
pressure setting the proGAV Check-mate has
to be put centrally over the valve. The proGAV
Check-mate will immediately start to move. If
it remains stable, the pressure setting can be
read in alignment to the inlet connector.
To adjust a new pressure setting, the proGAV
Check-mate has to be placed centrally over the
valve. The new pressure setting has to point
towards the proximal catheter (leading to the
ventricle). By pressing down slightly the proGAV
Check-mate, the brake of the valve is decou-
pled, the rotor turns and the opening pressure
of the proGAV 2.0 is changed.
Please be aware that the steps for changing
the pressure setting should not be more than
8 cmH
O per step.
2
Fig. 14: proGAV Check-mate
Caution: Due to magnets inside the
proGAV 2.0 Tools, do not use the pro-
GAV 2.0 Tools nearby pacemakers. Further
more do not use the proGAV 2.0 Tools near-
by MRI scanner, since ther is a danger of
damaging the MRI-scanner.
INSTRUCTIONS FOR USE |

ADJUSTING THE ADJUSTABLE DP-UNIT

Please verify specifically before using any tool
for verifying or adjusting the opening pressure:
For the adjustable DP-unit use either the
proGAV 2.0 Tools
a)
Fig. 15:
a) proGAV 2.0 Compass
b) proGAV 2.0 Adjustment Tool
and with a combination of proGAV 2.0 with an
adjustable gravitational-unit, use only proSA
Tools
Fig. 16: proSA Adjustment Tool
GB
b)
21