Input and output insulation
NOTE: test conditions: 4 kV rms ac for one minute.
Type
Power supply
(H or L)
Measurement
inputs
Digital outputs
Serial port
Ethernet port
Breakdown
• NA : combination not possible
• 4: 4 kV rms insulation (EN 61010-1, IEC 60664-1, overvoltage category III, pollution
degree 2, double insulation on system with maximum 300 Vrms grounding)
Environmental specifications
Operating temperature
Storage temperature
NOTE: R.H. < 90 % non-condensing @ 40 °C / 104 °F.
54
8021572 | WM20 | © 2016 | CARLO GAVAZZI Controls SpA
Power supply
Measurement
(H or L)
[kV]
-
4
4
4
4
Digital
inputs
outputs
[kV]
[kV]
4
4
-
4
4
-
4
4
4
4
From -25 to +55 °C/from -13 to +131 °F
From -30 to +70 °C/from –22 to +158 °F
Ethernet
Serial port
port
[kV]
[kV]
4
4
4
4
4
4
-
NP
NP
-