K-Patents PR-23 Refractometer System: Sensor verification
Sensor verification
Sensor information
Test results
Traceability and conformance
1 of 1
brinda luego de presionar
valores de CCD y TEMP se encuentran en la pantalla de resultados de la verificación;
consulte la Figura 13.10.
Page loaded on February 2, 2009 at 03:19:28 PM
Serial number:
R05676
Nominal nD
Measurement
at 25°C
at
Measured
Accuracy
actual
nD
temp.
1 1.3300
1.3297
1.3296
0.0001
2 1.3700
1.3697
1.3696
0.0001
3 1.4200
1.4196
1.4197
0.0000
4 1.4700
1.4697
1.4697
0.0000
5 1.5200
1.5196
1.5196
0.0000
Verification result: pass
Measurements should be done by using Cargille Refractive Index Liquids with
stated in accuracy of ±0.0002 R.I.
Cargille Index of Refraction Liquids are manufactured and calibrated using
instruments whose accuracy is verified by daily comparision to N.I.S.T. (N.B.S.)
traceable standards. N.I.S.T. standards are based on angle measurements using
a Wild divided circle spectrometers.
Figura 13.8 Certificado
de verificación del instrumento
Figura 13.9 Verificación fallida
(Verificar) en el paso 2 de la verificación. La lista de
VERIFY
Manual de Instrucciónes PR-23
http://192.168.23.250/verif.shtml
PROCESS REFRACTOMETER PR-23
Sensor
Result
Head
Head
Supply
temp.
hum.
current
°C
%
mA
pass
30
3
39
pass
30
3
39
pass
30
3
39
pass
30
3
39
pass
30
3
39
2.2.2009 15:21