 Connect the device.
1. After installation, wait for four hours before switching the centrifuge on in
order to avoid damage to the compressor.
2. Check that the mains voltage and frequency match the requirements on the
device name plate(See rear side of the device) and then connect the device
to the power supply.
 Remove the transport protection material from the rotor chamber.
1. Switch on the device at the mains power switch.
2. Open the centrifuge lid using the open button.
3. Remove the transport protection material.
4. Place the rotor vertically onto the motor shaft.
5. Turn the rotor nut using the rotor key clockwise until the rotor nut is tightened.
 The device is now ready to use.
Retain the packaging and all transport protection material for shipping the
device at a later date.
8.3.3

Packing

Pack the centrifuge in reverse order.
8.3.4

Passing on the Device