TM
Frontier
Centrifuge FC5515/FC5515R
13
6
3.4
Radius correction
If you use adapters or reducers it could change the centrifugal radius of the respective rotor. In that case you can
correct the radius manually. Please proceed as follows:
Close the lid, then press the key "Time" (6) (refer to figure 21) and the key "Prog/Setup" (11) at the same time
and hold them.
In the display "Time/Prog" (A-3) the word "radius" (M9) appears. By the adjusting knob (1) you can preselect the
respective radius correction (See Table 7, APPENDIX) in steps of 0.1 cm. As soon as you have set a radius
correction the word "radius" (M9) appears. This word will be visible until you put the radius correction back to 0
again.
3.5
Program
3.5.1
Storage of programs
You can store up to 99 runs with all relevant parameters, including the used rotors. You can use any free program
number and call it up again.
Put the needed rotor into the centrifuge. By pressing the key "Prog/Setup" (11) in the display "Time/Prog" (A-3)
the word "programm" appears. With the adjusting knob (1) you can chose the desired program number.
If a program number is already occupied, in the display "RPM | RCF" (A-1), the words "rotor" (M3) and "xx" (M4)
will appear. In case of free program numbers, 0 appears.
M9
A-4
Figure. 20
A-3
Figure. 21.
EN-11
10
11
1
6