Call Up Operating Data - OHAUS Frontier FC5515 Manual De Instrucciones

Ocultar thumbs Ver también para Frontier FC5515:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

TM
Frontier
Centrifuge FC5515/FC5515R
A-1
4
5
4.1.8

Call up operating data

ATTENTION:
This should only be performed by advance user or service engineer.
In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceed as
described under point 4.1.2 to enter this program mode. Press the key "Accel/Decel" (5). In the display "Acc/Dec"
(A-2) the word "Service" flashes.
With the adjusting knob (1) the different information can be accessed:
A= previous starts of the centrifuge
H= previous operating hours
S= software version
r= converter software
E= list of previous error messages
h= running time of the motor
The list of the last 99 error messages can be looked over by pressing the key "RPM | RCF" (4) and scroll through it
by the adjusting knob (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to
"Table 6: error messages" (See APPENDIX).
To change back to normal program mode again, switch off the centrifuge for a short period.
A-1
4
5
A-2
Figure. 36
A-2
Figure. 37
EN-19
1
1

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Frontier fc5515r

Tabla de contenido