Division Setting

LCD shows [d
x]. Then press
key to select 1, 2, 5, 10, 20, 50, and then
press
key to confirm and enter the decimal point setting.
Ex. When display shows [d
5]. Press
key, the division set as 5, then
enter the decimal point setting.

Decimal point setting

LCD shows [p x]. Then press
key to select 0, 0.0, 0.00, 0.000, 0.0000, P.00000,
and then press
key to confirm and enter the maximum capacity setting.
Ex. When display shows [p 0.000], press
key, the division set as 0.000,
then enter the maximum capacity setting.

Maximum capacity setting

LCD shows [ fULL ], around 2-3 seconds then LCD shows [000.000]. The
decimal point was set as 3 decimal points, then press
key to move the
flashing digits to the right in circle to select the digits that will increase
the number, then press
key to increase the number until the number
appears, then repeat the above moves until LCD shows the maximum
capacity then press
key to enter calibration setting.
Ex. LCD shows [50.000] then press
key to enter calibration setting.

Zero point calibration

LCD shows [CaL 0 ]. Do not put any objects on the pan when display shows
this, then scale will move to next calibration setting.
25