Configuration - Dräger DrugTest 5000 Instrucciones De Uso

Ocultar thumbs Ver también para DrugTest 5000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

After finishing the test:
Open the door of the Analyzer and remove the test cas-
sette with the cartridge attached from the lower com-
partment of the Analyzer – figure 6.
Close the door of the Analyzer.
– After displaying and, if necessary, printing the test result
the current analysis is completed and the unit is ready for
carrying out further tests.
– Before switching off and transporting the unit remove the
used test cassette.
Switching off the Analyzer
Keep button
OK
pressed. The device switches off after
approx. 3 seconds. To load the battery, the power cord
can be connected until the device stowed away, for ex-
ample for a transport.
Menu
The menu can be reached via the » Menu « item when the
device is ready for measurement.
Detailed information is included in the technical handbook
*)
Dräger DrugTest System
.
The user navigates through the menu via three buttons with
their respective functions displayed on the screen
(Softkey).
– If no buttons are pressed within 120 seconds the display
will automatically return to the measurement mode.
Navigation in the menu
Graphical symbols (icons) simplify the navigation through
the menu levels:
Return to the next higher menu level
Closed folder:
This item contains further functions or sub-menus
Open folder
The functions and sub-menus shown here are con-
tained in this item.
Function
If activated, functions can be executed in a single step
or several steps.
Selection activated
Press the »
« key for functions which can be selected
OK
and activated.
*) The technical handbook can be downloaded for free from the Dräger
homepage (www.draeger.com).
8

Configuration

The device can be connected to a PC using the integr ated
USB interface. This interface can be used by qualified
users and the Technical Service of Dräger Safety to set up
the device (e.g. for data storage or individual sample allo-
cation). For further information on this topic, please con-
tact Dräger Safety or a service organisation authorised by
Dräger Safety.
Fault – Cause – Remedy
The following table describes possible faults and system
messages.
If you encounter problems other than the ones listed in the
table while using the device or if your problem cannot be
solved by following the listed remedies, please contact the
Dräger Safety Service department or a service organisation
authorised by Dräger Safety (see section "Problem Solu-
tion and Repair" in chapter "Maintenance" on page 8).
Faults
Cause
No screen display
Battery totally dis-
within 10 seconds
charged and no
after powering up
external power
the device
supply estab-
lished
Audio alarm and
The door has
displayed mes-
been opened in
sage » Close
the analysis pro-
door «
cess
Audio alarm and
A Test Cassette is
displayed mes-
in the device dur-
sage
ing the automatic
» Remove Test
self check
Cassette «
Smoke / pungent
Incorrect battery
smell
charger used
Maintenance
The Dräger DrugTest 5000 Analyzer is a robust device which
requires minimum service.
Maintenance
We recommend to let the device get ser viced regularly
every 12 months.
The maintenance should be carried out by the Dräger
Safety Service department or a service organisation
authorized by Dräger Safety.
We recommend to enter a service contract.
Problem Solution and Repair
If you should encounter any problem which cannot be
solved with the suggestions given in table "Fault – Cause –
Remedy" on page 8, please inform the Dräger Safety Ser-
vice department or a service organisation authorised by
Dräger Safety.
Remedy
Connect battery
charger and
Cleaning
external power
supply.
If necessary, you can clean the outside of the device with
a soft cloth moistened with a mild soap solution or a com-
mercially available laboratory cleaning agent.
Close the door.
Do not dip the device into liquids and do not let any liquid
get into the terminals!
– The inside of the device does not require cleaning when
used as intended.
Open the door,
remove the Test
Cassette, close
Battery maintenance
the door.
The unit is equipped with an integrated lead storage bat-
tery.The unit consumes a small standby current even if
Use battery
switched off, which can discharge a fully charged battery in
charger supplied.
approx. 6 months. The capability of the batter y is retained
at an optimum and for the long term if prolonged periods of
Should the prob-
lem persist: Con-
low charge are avoided. To to so the following measures are
tat Dräger Safety
required:
– If the unit indicates a low battery charge, connect the
battery charger and completely charge the battery as
soon as possible.
– Prior to prolonged storage of the unit (several weeks)
charge the battery completely. Before the next use an-
other complete charge is recommended.
– For long-term storage of the unit (several months)
charge the battery completely every six months or keep
the battery charger connected throughout the duration
of storage.
– During storage at temperatures below –10
the battery charger connected throughout the duration
of storage.
o
C also keep

Publicidad

Tabla de contenido
loading

Tabla de contenido