Configuration - Dräger PIR 7000 Instrucciones De Uso

Ocultar thumbs Ver también para PIR 7000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 56
Action
Status Indicator
Display of the span deviation:
Wait approx. 1 to
Green status light
2 minutes, until the
flashes a single rhythm:
yellow status light
goes out. The
Green status light
flashing rhythm of the
flashes a double rhythm:
green status light
indicates the current
deviation.
Green status light
flashes a triple rhythm:
Carry out the span calibration:
Put the magnetic
Green and yellow
wand on the »S «
are on
mark and hold it there.
Remove the
Green / yellow flash
magnetic wand.
alternately
After reaching the gas concentration prior to calibration (accuracy: ±5 %), the gas transmitter
automatically terminates the calibration and switches back to normal operation (green signal
light on).
After finishing the calibration process or after exceeding the timeout value, shut down calibration
gas supply and remove the gas supply accessories used in the calibration process, if necessary.
The factory default calibration gas and concentration thereof are specified on the
configuration label.
If there is deviation from these settings, the modified parameters should be transferred using
the Remote Access Box Dräger RAB 7000 or using a computer with Dräger PolySoft
software or using a manual control unit made by HART
sure to put the modified parameters on the configuration label. The recommended calibration
gas concentration is between 40 % and 60 % of the full-scale deflection.
After changing the calibration gas or calibration gas concentration, the output value must be
checked after the initial application of the calibration gas (e.g. on the connected central device or
in the calibration screen when calibrating using a computer and the Dräger PolySoft software).
Errors / Complications during calibration
Action
Status Indicator
Yellow flashes
quickly
Yellow flashes
quickly and green
flashes approx. every
2 seconds
Output Signal Meaning
Maintenance
The deviation is less than the
signal
set "calibration range limit".
Maintenance
Slight deviation.
signal
Maintenance
The zero point deviation is
signal
larger than ±15 % of the
calibration gas concentration.
(Timeout after approx. 15 min)
Maintenance
Adjustment is acknowledged.
signal
Maintenance
Span calibration is terminated.
signal
WARNING
®
and checked after being read in. Be
Output Signal
Meaning
Maintenance
Device has an error or has
signal
detected complications.
Span calibration not possible
because the last zero point
calibration was carried out
more than 24 hours ago. Zero
point calibration required.
Action
Status Indicator
Put the magnetic
Green and yellow are
wand on the » 0 «
on
mark for zero-point
calibration or on the
» S « mark for span
calibration and hold
it there.
Remove the
Green / yellow flash
magnetic wand.
alternately and slowly
The gas transmitter automatically terminates the calibration, then switches back to normal
operation (green signal light on).
After cancelling the calibration process or after exceeding the timeout value, shut down gas
supply and remove the gas supply accessories used in the calibration process, if necessary.
Terminating the calibration
Action
Status Indicator
Put the magnetic
Yellow flashes
wand on the » S «
quickly (for approx.
mark for zero-point
2 sec.)
calibration or on the
» 0 « mark for span
calibration and hold
it there.
Green and yellow
are on
Remove the
Green / yellow flash
magnetic wand.
alternately and slowly
The gas transmitter automatically terminates the calibration, then switches back to normal
operation (green signal light on).
After cancelling the calibration process or after exceeding the timeout value, shut down gas
supply and remove the gas supply accessories used in the calibration process, if necessary.
Configuration of the gas transmitter
To customise the configuration on a device with standard configuration, use a computer and
the Dräger PolySoft software (see Instructions for Use for the PC software). After making the
changes to the configuration, reread the configuration and check it.
In delivery state, the following configuration is set (provided that this is not a built-to-order unit):

Configuration:

Type 334
Conversion table %LEL
Measured gas
unit
measuring range
0 ... 100 %LEL
Calibration gas
unit
Calibration gas concentration
Maintenance signal
Fault signal
Beam block warning (inactive)
Output Signal
Meaning
Maintenance
Error display message is
signal
acknowledged.
Maintenance
The calibration is cancelled
signal
without saving the values.
Output Signal
Meaning
Maintenance
Device has detected an
signal
operator cancellation.
Maintenance
Device acknowledges the
signal
termination.
Maintenance
The calibration is cancelled
signal
without saving the values.
Dräger PIR 7000
Dräger
Type 340
PIR 7200
Category 1
based on NIOSH
Methane
Propane
Carbon dioxide
%LEL
%LEL
0 ... 100 %LEL
0 ... 10 % vol.
Methane
Propane
Carbon dioxide
%LEL
%LEL
50 %LEL
4 vol.%
3 mA
<1.2 mA
2 mA
– – –
vol.%
vol.%
27