Guidance and manufacturer's declaration - electromagnetic emissions
Emissions test
RF emissions CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic emissions
Immunity test
IEC60601-1-2 test level
Surge
±1kV differential mode
IEC 61000-4-5
±2kV common mode
Voltage
Voltage reduction in
dips, short
% of UN: 100% for 1
interruptions
period, 100% for 0.5
and voltage
periods, 30% for 25
variations on
periods, 100% for 250
power supply
periods
input lines IEC
61000-4-11
18
Compliance
Electromagnetic environment - guidance
Group 1
This equipment uses RF energy only for its internal functions. There-
fore its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
Class A
This equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
Complies
domestic purposes.
Compliance
±1kV differential mode
±2kV common mode
Voltage reduction in
% of UN: 100% for 1
period, 100% for 0.5
periods, 30% for 25
periods, 100% for 250
periods
Electromagnetic environment - guidance
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the MiniLift160Classic requires continued
operation during power mains interruptions, it
is recommended that the MiniLift160Classic be
powered from an uninterruptible power supply
or a battery.
U S E R M A N U A L