Filling a Cylinder
Once the cylinder is properly connected, you can start the fill process.
1.
Press the stop/start button.
2.
As the oxygen cylinder fills, the level indicators on the UltraFill will illuminate. The first level indicator blinks while
the oxygen cylinder is filling, then turns solid when the device is 25% full. As the oxygen cylinder continues to fill,
each level indicator will blink, then illuminate.
3.
When the cylinder is full, all level indicators will be solid. The UltraFill will automatically turn off and the cylinder
can be disconnected.
Note: the cylinder can be removed before it is completely full. Simply follow the steps outlined in the section called
"Disconnecting Cylinder".
Note: if the yellow LED is lit continuously and the red LED is flashing with a constant audible alarm, the O
cylinder has dropped below 90%. Remove and empty the cylinder (see the sections of this manual called "Disconnecting
a Cylinder and "Emptying a Cylinder"). Refer to the "Troubleshooting Guide" in this manual for more information. If the
problem persists, contact your homecare provider.
Cylinder Size
2000 psi (135 bar)
M4
M6
M9
MD15
ME24
3000 psi (200 bar)
MB08
MC13
MD22
ME36
UltraFill User Manual
Volume (L)
103
148
246
413
679
232
354
604
992
Typical Fill Times
45 min
1 hr 5 min
1 hr 50 min
3 hr 5 min
5 hr
1 hr 45 min
2 hr 35 min
4 hr 25 min
7 hr 20 min