intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an
environment.
e
t
mISSIonS
eSt
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
g
m
uIdance and
anuFacturer
the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
I
t
mmunIty
eSt
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical Fast Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
UltraFill User Manual
'
d
S
eclaratIon
c
omplIance
Group 1
Class B
NA
NA
'
d
S
eclaratIon
Iec 60601
t
l
eSt
evel
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV Line to Line
<5% U
T
(>95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
) for 5 cycles
T
70% U
T
(30% dip in U
) for 25 cycles
T
<5% U
T
(>95% dip in U
) for 5 sec
T
3 A/m
– e
lectromagnetIc
e
lectromagnetIc
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
No test requirements for devices that operate at 115 VAC 60 Hz.
– e
I
lectromagnetIc
c
l
omplIance
evel
±6 kV contact
±8 kV air
±2 kV for power supply mains
±1 kV line to line
<5% U
T
(>95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
) for 5 cycles
T
70% U
T
(30% dip in U
) for 25 cycles
T
<5% U
T
(>95% dip in U
) for 5 sec
T
3 A/m
e
:
This device is
mISSIonS
e
nvIronment
:
This device is intended for use in
mmunIty
e
lectromagnetIc
– g
uIdance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
home or hospital environment.
Mains power quality should be that of a typical
home or hospital environment.
Mains power quality should be that of a typical
home or hospital environment. If the user of
the device requires continued operation during
power mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical hospital or home environment.
– g
uIdance
e
nvIronment
25