Troubleshooting Guide - Respironics UltraFill Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

t
roubleSHootIng
UltraFill Station
Problem
Power switch is on but no
LED's are illuminated.
Cylinder is connected
but no blue LED's are
illuminated.
Cylinder is connected,
at least one blue LED is
illuminated, but fill will not
start.
Yellow LED is blinking
with periodic beep.
Yellow LED is illuminated.
UltraFill User Manual
g
uIde
Possible Cause
Device is not plugged in or
there is a power failure.
Cylinder is not fully
connected.
Cylinder is not fully
connected.
Internal part failure.
Low oxygen pressure.
Cylinder valve may be
open. (Only applicable if
the OCD/Regulator is not
attached).
The UltraFill device has
detected low oxygen purity.
Your Action
Check the power outlet and verify that the device is
plugged in.
Press cylinder disconnect button and remove
cylinder. Verify there is no obstruction to cylinder
fill connection. Retry connecting cylinder with firm
downward force.
DO NOT LUBRICATE CYLINDER FITTING WITH
OIL OR GREASE.
If the condition persists contact your home care
provider.
Press the cylinder disconnect button and remove the
cylinder. Verify there is no obstruction to the cylinder
fill connection. Retry connecting the cylinder with
firm downward force.
DO NOT LUBRICATE THE CYLINDER FITTING
WITH OIL OR GREASE.
If the problem persists, contact your home care
provider.
Press the start/stop button. Verify that the oxygen
concentrator and UltraFill are connected properly.
Check for kinked or obstructed tubing. If the OCD/
regulator is not attached to the cylinder, verify that
the cylinder valve is closed.
If the problem persists, contact your home care
provider.
Press the start/stop button.
Check the flow meter on the concentrator and verify
the flow is ≤ 3 LPM (5L concentrator) or ≤ 7 LPM
(10L concentrator).
Wait 5 minutes and press the start/stop button.
If the problem persists, contact your home care
provider.
19