Cylinder Use Times
The pulse dose will vary on each individual's breathing patterns. The use time will vary depending on the patient's
prescription, breath rate and type of conserving device being used. All use times are calculated using a setting of 2 and
at a breath rate of 20-breaths per minute for pulse dose.
Respironics will offer both an oxygen conserving device (CHAD Therapeutics Bonsai) and two regulators (CHAD
Therapeutics 4800 and 3100) with our system package. The use times are calculated below:
Volume
(L)
Cylinder
Type
M4
103
M6
148
M9
246
MD15
413
ME24
679
MB08
232
MC13
354
MD22
604
ME36
992
UltraFill User Manual
1
2
6 hr
3 hr
26 min
51 min
9 hr
5 hr
15 min
33 min
15 hr
9 hr
23 min
14 min
25 hr
15 hr
48 min
29 min
42 hr
25 hr
26 min
28 min
14 hr
8 hr
30 min
42 min
22 hr
13 hr
8 min
17 min
37 hr
22 hr
45 min
39 min
62 hr
37 hr
12 min
Duration (at Bonsai Settings)
3
4
2 hr
2 hr
45 min
9 min
3 hr
3 hr
57 min
5 min
6 hr
5 hr
34 min
8 min
11 hr
8 hr
1 min
36 min
18 hr
14 hr
6 min
9 min
6 hr
4 hr
11 min
50 min
9 hr
7 hr
26 min
23 min
16 hr
12 hr
6 min
35 min
26 hr
20 hr
27 min
40 min
5
6
1 hr
1 hr
53 min
43 min
2 hr
2 hr
43 min
28 min
4 hr
4 hr
31 min
6 min
7 hr
6 hr
34 min
53 min
12 hr
11 hr
27 min
19 min
4 hr
3 hr
15 min
52 min
6 hr
5 hr
29 min
54 min
11 hr
10 hr
4 min
4 min
18 hr
16 hr
11 min
32 min
7
2 lpm
CF
1 hr
52 min
27 min
2 hr
1 hr
6 min
14 min
3 hr
2 hr
29 min
3 min
5 hr
3 hr
51 min
27 min
9 hr
5 hr
37 min
40 min
3 hr
1 hr
17 min
56 min
5 hr
2 hr
1 min
57 min
8 hr
5 hr
33 min
2 min
14 hr
8 hr
3 min
16 min