3.

GETTING STARTED

3.1.

Connecting the test leads

Connect the black lead to the COM socket (for all measurements). Depending on the
position of the selector switch, connect the red lead as follows:
V
, V
LOW Z
AC+DC
500 mA
(500 mA
DC
10 A
(10 A
DC
3.2.

Switching on the instrument

The selector switch is on the OFF position.
Turn the selector switch to the required function.
All segments of the display come on for a few seconds. The instrument is then ready for
measurements.
3.3.

Switching off the instrument

The instrument can be switched off manually by returning the selector switch to the OFF
position, or automatically after approximately half an hour if no key is pressed or the
switch is not operated.
F
Note
For user safety, automatic shutdown is also disabled when a measured
magnitude (voltage/current) present at the input exceeds dangerous level
(
3.4.

Special configuration

To adapt the configuration of the instrument to the measurement environment, the user
can choose 50 Hz or 60 Hz rejection :
Switch on with the rotary switch while holding down the MEM key. The selection is
reversed from the last configuration, is displayed for two seconds and remains backed
up in non-volatile memory.
3.5.

Multimeter maintenance

3.5.1. Fuse self-test

This check can be carried out with the multimeter in use without opening up the casing.
- 0.63 A fuse (F1) : set the rotary switch to the
and VΩ sockets. The display should read approx. 0.001 V. If the digital display indicates
an overflow « .OL ». The fuse is blown.
- 10 A fuse (F2) : set the rotary switch to the
Ω sockets. The display should read approx. 0.001 V. If the digital display indicates an
overflow « .OL ». The fuse is blown.
5000-count portable digital multimeter, with current input
Rotary selector switch position
, V
, ADP
(ADP
DC
DC
)
AC+DC
)
AC+DC
indicator displayed).
, Ω,
), FREQ,
AC+DC
position and short-circuit the 500 mA
position and short-circuit the COM and V
Input terminal
V Ω
,
500 mA
10 A