6. Operation
6-1 Programming the Micromotor Operation
The Control Unit can memorize 8 programs. Each program includes the following functions which will be
automatically performed when the appropriate program number is selected.
Gear ratio of contra angle handpieces
Speed
Direction of rotation
Torque upper limit
Coolant Solution Flow
Light Level
1)
Turn on the power by pushing the Main Power Switch toward [-]; on
power up IMPLANT SYSTEM Number 1 program number 1 is displayed by default.
2)
Select a program number by using either step (a) or step (b):
(a) Press the PROGRAM (+ -) Key on the Control Panel until the desired program number is displayed.
(b) Press the PRG(Program) Button on the Foot Control until the desired program number is displayed.
3)
Selecting the Rotate direction of the handpiece. Select it by using either step (a) or (b):
(a) Press FWD/REV Key on the Control Panel until the desired direction is displayed.
(b) Press the Forward/Reverse Button on the Foot Control until the desired direction is displayed.
NOTICE
• There will be a beeping sound while "R" is pressed.
• Beeping sounds heard during rotations and when the device is stopped are not the same. Different sounds do not
indicate a malfunction.
4)
Selecting the Gear ratio of contra angle handpieces relevant to the program; Press the GEAR Key to select the
gear ratio of the handpiece (Gear Ratio will display on the LCD).
5)
Set the required max operating speed by pressing the SPEED (+ -) Key. Each time this Key is pressed display
changes to the next speed level. By pressing this Key for more than 1 second brings the speed quickly to the
next level until the speed display reaches its upper or lower limit.
- When the speed setting reaches the upper or the lower limit, an audible beep is heard and the speed setting
cannot be changed any further.
6)
Set the torque upper limit by pressing the TORQUE Key on the Control Panel. Each time this Key is pressed
display changes to the next torque level. By pressing this Key for more than 1 second brings the torque quickly
to the next level until the torque display reaches its upper or lower limit.
- When the torque setting reaches the upper or the lower limit, an audible beep is heard and the torque setting
cannot be changed any further.
7)
The level of Coolant Solution Flow volume has 6 Flow level (0-5) (0 = no coolant Flow). Select a Coolant Solution
Flow Volume using either step (a) or step (b):
(a) Press the COOLANT Key on the Control Panel until the desired flow level is displayed.
(b) Press the Coolant Solution Flow Volume Button on the Foot Control until the desired flow level is displayed.
8)
After completing steps 2) - 7). press and hold MEMORY Key until beep is heard. The beep confirms that the
programming is completed. If you hear a short beep when the MEMORY Key is first pressed ignore this sound
and keep the MEMORY Key depressed until a long beep is heard.
* Repeat the above steps 2) - 8) to program any one of the 8 available programs.
6-1-1 Program Skip Function
When switching programs using the Foot Control, Program Numbers that are not chosen can be chosen ahead of time.
1)
Choose Program Numbers that you want the Control Unit to skip.
2)
Use the SPEED (-) Key to display the Minimum Rotation Speed.
3)
Release the SPEED (-) Key, and then push it again. Display "---" on the LCD.
4)
Push the MEMORY Key until it beeps (about 2 seconds) to memorize.