Handling Of Exhausted Batteries; Device Alarm; Changing The Sensor; Trouble Shooting Warnings/Errors - Dräger Pac 3500 Instrucciones De Uso

Ocultar thumbs Ver también para Pac 3500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24

7.1 Handling of exhausted batteries

Danger of explosion!
Do not throw used batteries into fire or try to open them by force.
Dispose of the batteries in accordance with local regulations.
Spent batteries may be returned to Dräger for disposal.
Usable life alarm (only valid for Pac 3500)
– Before the end of the device's usable life, a warning period begins. During
this period the remaining life time flashes just after turning the device on, e.
g. "30" / "d".
– To acknowledge this message [OK] must be pressed. After that, the device
can be used normally.
– After the usable life has expired the text "0" / "d" will alternate in the display
and cannot be acknowledged. Dräger Pac 3500 will not longer measure and
may be returned to Dräger for recycling or disposal.

7.2 Device alarm

– The alarm and LED will be activated three times, periodically.
– The [X] icon is flashing, a 3 digit error code will be shown in the display.
– If an error appears in the display see section 9 and if necessary please con-
tact Dräger.

8 Changing the sensor

Danger of explosion! Do not change the sensor in explosion-hazard areas.
Substitution of components may impair intrinsic safety!
To prevent ignition of flammable or combustible atmospheres, and to avoid
compromising intrinsic safety of the equipment, read, understand, and
adhere to the maintenance procedures below.
Take care when changing the sensors not to damage or short out compo-
nents, and do not use sharp tools to remove the sensors.
Damage to components!
The instrument contains components that are sensitive to electrical charges.
To prevent damage to the instrument, make sure that the person working on
the instrument is earthed prior to opening it for sensor replacement. Earth
contact can, for example, be ensured by an ESD (electrostatic discharge)
workplace.
Replace sensor when instrument can no longer be calibrated!
6
WARNING
WARNING
CAUTION
Notice
Use only the DrägerSensor XXS of the same gas type!
– Turn the instrument off.
– Unscrew the 4 screws from the back case.
– Open the front case and remove the battery.
– Remove the sensor.
– Insert the new sensor.
– Insert the battery according to specified polarity (+/–).
– Place front case back and fasten it by tightening the 4 screws of the back
case.
– After inserting the battery a sensor warm up time of around 15 minutes is
needed.
– Countdown in seconds is shown until warmup time has passed.
– After changing the sensor and after the warm up time is finished the
instrument must be calibrated.

9 Trouble shooting warnings/errors

Error
Cause
code
010
"Horn alarm elements" X-dock
test failed
011
"LED alarm elements" X-dock test
failed
012
"Vibration alarm elements" X-dock
test failed
013
Parameter check failed
014
Instrument locked by X-dock
100
Flash/EEprom write failed
104
Flash check sum wrong
105
Broken or missing O
sensor
2
106
Most recent settings restored
107
Self test failed
108
Configuration not up-to-date
109
Configuration error
161
Set operating time of the instru-
ment expired
Notice
Remedies
Repair if necessary and test again
with X-dock
Repair if necessary and test again
with X-dock
Repair if necessary and test again
with X-dock
Correct parameters and test again
with X-dock
Cancel X-dock lock
Contact Service
Contact Service
Replace O
sensor
2
Check settings and recalibrate
instrument
Contact Service
Reconfigure with the latest Dräger
CC-Vision software version
Reconfigure instrument
Set new operating time of the
instrument

Publicidad

Tabla de contenido
loading

Tabla de contenido