1. Place the nut and metal seal of
the cable gland onto the shielded
cable.
4. Insert the grounding ferrule over
the shielding.
7. Use the crimping pliers to attach
the protective plugs to the wires.
Connecting peripherals
To conform with security standards and to respect the EMC compliancy of the analyzer, this procedure must be
followed for connecting all peripherals to the electronic unit of the analyzer.
10 English
2. Remove 8 cm of the outer
cabling to reveal the braided
shielding.
5. Use the clamp to fix the ferrule
securely to the cable.
8. Bring the metal shield up over
the ferrule and lock the two
together making sure the ferrule is
not exposed at the cable end of
the shield. Bring up the nut to
cover the shield.
W A R N I N G
3. Remove all but 12 mm of the
shielding to reveal the interior
wires, then roll the remainder of
the shielding back over the cable.
6. Remove 5 mm of the external
insulation of the wires.
9. Pass the threaded part of the
cable gland over the wires and
screw into the nut on the cable.
The cable is now ready to be
connected following the procedure
described in
Connecting
peripherals
on page 10.