Set Up; Operation Instructions; Programming Auto-Off Time; Maintenance - Health O Meter 594KL Instrucciones De Montaje Y Uso

Balanza electronica con silla
Ocultar thumbs Ver también para 594KL:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SET UP

Your scale is a precision instrument and should be handled accordingly. The Health o Meter Professional Digital Chair Scale is shipped completely
assembled, so no additional assembly required.
Some protective plastic, cardboard and foam will need to be removed before using.
1. Remove metal bands attached to small skid.
2. Remove cardboard at the front and the back end of the chair.
3. Remove protective foam and plastic cover on the display.
4. Plug the scale‟s AC adapter into the chair scale display and into the power source.
OPERATING INSTRUCTIONS
1. Make sure the scale is on a hard, level surface before using.
2. Press the ON / OFF button to turn on the scale.
3. The indicator will start the initialization count from 000000 to 999999.
4. Wait until „0.0‟ and the arrows above ZERO and STABLE appear on the display.
5. The model 594KL model is designed to provide convenient and accurate weighing to the less mobile. The side-arms and footrest are movable for
easy access and patient comfort (See Figure).
6. Place the patient on the scale. The scale will display the patient weight. (NOTE: the weigh will change until the patient is stable). Once the patient
is stable an arrow will appear above STABLE on the display.
7. Press and hold the KG / LB button for 5 seconds to select the weigh mode, either kilograms or pounds.
8. Weight may remain displayed on the scale by pressing the HOLD / RELEASE button. To release, press the HOLD / RELEASE button again.
NOTE: The scale will always default to the settings and units last used.
NO.
DESCRIPTION
1.
HOLD / RELEASE
2.
BACK LIGHT
3.
KG / LB
4.
TARE / ZERO
5.
ON / OFF

PROGRAMMING AUTO-OFF TIME

The 594KL has an Auto-Off feature that will automatically turn the scale off after a period of inactivity. This battery saving feature is programmable by
the user. The 594KL can be programmed to automatically power off from 0 (disabled) to 18 minutes. The default shut off value set by the factory is
2 minutes.
1. Press ON / OFF and wait for the display to show "0.0".
2. At the same time press both the BACKLIGHT and TARE / ZERO buttons. The display will show "t" along with a number representing the current
shut-off setting.
3. Use the BACKLIGHT button to change the shut off setting. The shut-off values are set in 2 minute increments. (" t 0" = disabled; "t 1" = 2 minutes;
"t 2" = 4 minutes ........ "t 9" = 18 minutes.)
4. Press HOLD RELEASE to save the desired turn off value. "CAL" will appear on the display.
5. Press BACKLIGHT to exit. Wait for "0.0" to appear and the scale is now ready for normal operation.
GENERAL
The Maintenance and Troubleshooting sections provide instructions for maintenance, cleaning, troubleshooting and operator replaceable parts for the
Physical Digital Chair Scale. Maintenance operations other than those described in these sections should be performed by qualified service personnel.

MAINTENANCE

Before first-use and after periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, refer to quali-fied
service personnel.
1. Check overall appearance of the total scale for any obvious damage, abuse, etc.
2. Inspect condition of AC adapter cord and prongs, replace if needed.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Disconnect the scale from the AC power source
1. Clean all external surfaces with a clean damp cloth or tissue. Mild soap and water solution may be used. Dry with a clean soft cloth.
2. Do not immerse the scale into cleaning or other liquid solution.
3. Do not use Isopropyl Alcohol or other solutions to clean the display surface.
FUNCTION
Holds the value of the weighed object on the
display until the button is pressed again to
Turns display back light on and off
Pressing and holding button for 5 seconds
toggles between kilograms or pounds.
Zeros the scale prior to weighing or sets a
Turns scale ON and OFF
3

Publicidad

Tabla de contenido
loading

Tabla de contenido