Hach 8310 Manual De Usuario página 6

Ocultar thumbs Ver también para 8310:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
• B: Good - satisfactory immersion of the electrode surfaces and good sample flow.
• C: Poor - incomplete immersion of the electrodes or incorrect sample flow.
On piping
Immerse the internal electrode completely in the process sample.
Note: In the following illustrations, the arrows indicate the sample flow direction.
Figure 2 Model 8315 sensor mounting example
Figure 3 Model 8394 sensor mounting example
The 8394 sensor installs perfectly in a Tri-Clover
(A), and also at a 90° angle starting from 2 inches (B). All Tri-Clamp
3A standards for CIP (cleaning In place).
In a bypass
Polymetron flow chambers are designed not to retain air bubbles. To facilitate the extraction of
bubbles, use a minimum flow rate of 20 L/h (ideally 60 L/h).
Note: The progressive accumulation of bubbles on the surface of the sensor reduces the active surface, increases
the cell constant, and leads to an abnormally low conductivity measurement.
6 English
®
Tri-Clamp
Tee starting from 1.5 inch diameter
Tees are in compliance with

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

831183128315831683178394

Tabla de contenido