Ottobock 2R67 Instrucciones De Uso página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
► If the product and its components have been subjected to extreme
loads (e.g. due to a fall, etc.), then the product must be inspected for
damage immediately. If necessary, forward the product to an author­
ised Ottobock Service Centre.
CAUTION
Remaining in areas outside the allowable temperature range
Falling due to malfunction or the breakage of load-bearing product com­
ponents.
► Avoid remaining in areas with temperatures outside of the permissible
range (see Page 22).
CAUTION
Use of unsuitable prosthetic components
Falling due to unexpected behaviour of the product or breakage of load-
bearing components.
► Use the product only in combination with components approved by
Ottobock (see instructions for use and catalogues).
CAUTION
Improper use
Injury due to faulty operation or malfunction of the product.
► Instruct the patient in the proper use of the product.
CAUTION
Penetration of dirt and humidity into the product
>
Falling due to unexpected product behaviour as the result of malfunc­
tion.
>
Falling due to breakage of load-bearing components.
► Ensure that neither solid particles, foreign objects nor liquids penetrate
into the product.
► Should the product come into contact with liquids, allow the product to
dry. The product must be inspected by an authorised Ottobock Service
Centre.
► Should the product come into contact with salt water, immediately
clean it with a cloth moistened with fresh water and let it dry. The
product must be inspected by an authorised Ottobock Service Centre.
Ottobock | 17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4r394r404r854r86

Tabla de contenido