Safety Instructions - Dräger X-am 5600 Instrucciones De Uso

Ocultar thumbs Ver también para X-am 5600:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I,
Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C, and for areas where
gases or dusts of groups A, B, C, D and temperature class T3 or T4 may be present
(depending on the rechargeable battery and batteries).

Safety Instructions

To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following Caution and Warning statements:
Increased hydrogen concentrations within the measuring range of the Dräger
Sensor XXS H
HC may result in false alarms due to the additive effect on the
2
Dräger sensors XXS H
S, and XXS CO, XXS H
2
well as due to the negative effect on the Dräger Sensor XXS O
Only use power packs ABT 0100 (83 22 237), HBT 0000 (83 18 704) or
HBT 0100 (83 22 244). See marking on power pack for approved batteries and
related temperature class.
Do not change batteries in hazardous locations.
To reduce the danger of explosion, do not mix new batteries with old batteries
and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any
maintenance operations.
Substitution of components may impair intrinsic safety.
Not tested in an oxygen-enriched atmosphere (>21 % O
18
WARNING
S (LC) and XXS CO (LC) as
2
.
2
WARNING
WARNING
WARNING
WARNING
WARNING
CAUTION
).
2
What is What
1
2
2
3
8
+H
2
X-am 5600
7
6
4
0
5
1
Gas entry
2
Alarm LED
3
Horn
4
[OK] key
5
Power pack
6
[+] key
7
Display
8
Display sticker
H
added signal (optional)
2
Special symbols:
Fault message
Warning
Display peak value
Display TWA
Show STEL
Bump test mode
Fresh-gas adjustment
CO
2
Vol%
9
ch
4
%UEG
10
O
2
Vol%
CO
ppm
H
S
2
ppm
11
NH
3
ppm
12
13
14
15
9
IR interface
10
Fastening clip
11
Model plate
12
Charging contacts
13
Measured gas display
14
Measuring value display
15
Special symbols
1-button adjustment
Input gas adjustment
Password necessary
Battery 100 % full
Battery 2/3 full
Battery 1/3 full
Battery empty

Publicidad