Replacing the batteries / rechargeable batteries

Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from
different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
1)
Varta Type 4106
(power one) or
1)
Varta Type 4006
(industrial)
Alkaline batteries – T4 – (non rechargeable!)
1)
Duracell Procell MN1500
NiMH rechargeable batteries – T3 – (rechargeable)
1)
GP 180AAHC
(1800 mAh) max. 40 °C ambient temperature.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100)
with the appropriate Dräger charger. Charge NiMH single cells for ABT 0100
battery holder as directed by the manufacturer. Ambient temperature during
charging: 0 to +40 °C.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
Switching off the instrument:
Press and hold the [OK] and [+] keys simultaneously.
Loosen the screw on the power pack and remove the power pack.
24
WARNING
Battery holder (order no. 83 22 237):
Explosion hazard!
The Dräger X-am 5600 is only permitted to be operated with battery holder
ABT 0100 (X-am 5600), identified by the silver sticker.
Replace alkaline batteries or NiMHy rechargeable batteries. Ensure correct
polarity.
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
replace complete power pack.
Insert the power pack into the instrument and tighten the screw, the instrument
switches on automatically.
Charge instrument with NiMH power pack T4
(Type HBT 0000) / T4 HC (Type HBT 0100)
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the regulations for fire damp
and explosion protection.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100)
with the appropriate Dräger charger. Ambient temperature during charging:
0 to +40 °C.
Even if the instrument is not used, we recommend that you store the instrument
in the charger!
Insert the switched off instrument into the charger.
Display LED on the charger:
To maintain the lifetime of the batteries, charging is temperature controlled and
only performed in a temperature range of 5 to 35
range is left, the charging process is automatically interrupted and automatically
continued after the temperature range has been reached again. The charging
time is typically 4 hours. A new NiMH power pack reaches its full capacity after
three complete charge / discharge cycles. Never store the instrument for
extended periods without being connected to a power source (maximum of 2
months) because the internal buffer battery will drain.
WARNING
WARNING
Charge
Malfunction