Station Self-Test; Conducting A Test - Dräger X-dock 5300 Instrucciones De Uso

Ocultar thumbs Ver también para X-dock 5300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
5.3

Station self-test

A self-test is carried out:
When the station is started up.
When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
5.4

Conducting a test

WARNING
!
When using methane, propane or butane in the range
>100 %LEL, an exhaust hose (max. 10 m long) must
be connected to the exhaust outlet to ensure the
extraction of excess explosive gas.
NOTICE
i
i
Single mode is activated by default.
Several tests can be started and executed in parallel in
the Single mode.
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A testing of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
The following tests are preconfigured:
Quick bump test including checking the alarm
Test 1: QT
elements.
Extended bump test incl. zero-point check and
Test 2: EXT
checking the alarm elements.
Test 3: CAL
Calibration and checking the alarm elements.
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Conduct a visual inspection of the instruments (see
chapter 5.1 on page 22).
4. Insert the instruments into the modules (see chapter 5.2 on
page 22).
If Single mode is activated:
The default test is started automatically when the module
flap closes.
The Status LED flashes blue.
The individual test phases are displayed.
If Favorites mode is activated:
Select the desired test from favourites bar.
Test will be started automatically.
The status LED flashes blue.
The individual test phases are displayed.
Dräger X-dock 5300/6300/6600
If Test schedulermode is activated:
If necessary, log user out on the station (see chapter 4.5 on
page 20).
The preset test is performed according to the configured
time schedule.
If Log-in mode is activated:
Log user in on the station (see chapter 4.5 on page 20).
Select the desired test from favourites bar.
Test will be started automatically.
The status LED flashes blue.
The individual test phases are displayed.
Test passed:
Confirmation is shown on the display.
The Status LED flashes green.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
WARNING
!
Prior to each use of the gas detector, check whether it
is switched on and in measuring mode after removing
it from the X-dock. Otherwise the user may carry a
device that is switched off!
Use
0 0
01033286.eps
23