Conducting A Visual Inspection; Inserting Or Removing The Gas Measuring Device Into Or From A Module - Dräger X-dock 5300 Instrucciones De Uso

Ocultar thumbs Ver también para X-dock 5300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
Use
WARNING
!
If this function is deactivated, the corresponding
channel is not tested or calibrated.
To select the test gas for PAC 7000 OV:
1. Select PAC 7000 OV.
2. Select the required test gas from the list.
The following selection is available:
Ethylene oxide - EO (default setting)
Carbon monoxide - CO
3. Confirm selection with OK.
To set the gas cylinder level indicator:
NOTICE
i
i
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no..
1. Select Gas level monit..
2. Activate or deactivate check box Gas level monit..
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
> Station gas configuration.
3. Select desired gas inlet.
4. Select Further options and select
cylinder level indicator.
5
Use
WARNING
!
A defective pressure reducer on the gas cylinder can
lead to increased pressure in the station. The gas
hoses may loosen as a result and gas may escape.
Health hazard! Test gas must not be inhaled. Observe
the hazard warnings in the relevant Safety Data
Sheets.
Provide venting into a fume cupboard
or outside the building.
NOTICE
i
i
To prevent loss of gas, Dräger recommends closing
the gas cylinders when the station is left unattended for
long periods.
Adjustment may not be possible due to instrument and
channel errors.
22
5.1
A visual inspection of the instruments must be conducted
every time before being inserted into the station.
1. Check that the housing, external filters and the nameplates
are intact.
2. Check the battery contacts and sensor inputs for dirt.
i
i
5.2
to reset the gas
To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module.
3. Close the module flap.
The instrument will be detected automatically.
X-am-125+ module with charging function only:
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.

Conducting a visual inspection

NOTICE
Instruments that did not pass visual inspection must
not be inserted into the station. Otherwise the test
cannot be correctly assessed in its entirety.
Inserting or removing the gas measuring
device into or from a module
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx. 15
mintues after the last test.
Dräger X-dock 5300/6300/6600
2
1
00233286.eps

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X-dock 6300X-dock 6600

Tabla de contenido