PART C – TROUBLESHOOTING
The following table will enable a user to diagnose basic problems and implement a solution
that may enable the device to resume operation. If a probe is present and you need to
retrieve it urgently, see section B5.3.
If the problem persists, contact your Nanosonics representative
Symptom
1. There is no power to the
device
2. The screen is blank
3. The chamber door will
not open
4. The chamber door will
not close
5. The cartridge door does
not open
6. The cartridge door will
not close
7. The probe will not sit
correctly in the chamber
8. The cycle will not start
9. The device is beeping
10. Liquid is leaking from the
device
11. The chamber door is
®
trophon
EPR User Manual – N00020
Check for the following
The device is plugged in
The correct power cable for the region is used
The power is switched ON at the wall outlet
The power switch on the device is set to ON
There is power to the device
Follow screen messages for instructions
Any cycle (purge or disinfect) is complete
Chamber door handle can be completely opened
Follow screen messages for instructions
The probe is validated for use in the device
The probe Is loaded correctly
Door lock position: if in the 'locked' position, the door
will not close. To check, try opening the handle as if
opening the chamber. If the handle cannot be turned,
the door is in the locked position – see section B5.3
There is power to the device
Follow screen messages for instructions
The cycle (purge or disinfecting) is complete
The waste container is empty and fully inserted
Follow screen messages for instructions
Correct cartridge type: trophon NanoNebulant
Cartridge is correctly positioned
Cartridge lid has been removed
The probe is validated for use in the trophon EPR
The probe is loaded correctly
There is power to the device
Follow screen messages for instructions
The probe is loaded correctly
The chamber door is closed
If at the 'Clean and dry probe screen' and the YES
button will not register, contact your customer service
representative
If the screen reads READY and the start button will not
respond, contact your customer service representative
Follow screen message for instructions
Probe inside the device which needs to be removed
Warning: Any fluid leaking from the device may contain
hydrogen peroxide. If liquid or mist is seen coming from the
device at any time:
Do not come into contact with the mist or liquid
Wear appropriate PPE
Ensure area is well ventilated
Allow the device to complete the cycle
Turn off the device and remove the power cord
Contact your customer service representative
If the device is in sleep mode, select RESTART on the LCD