• If you press the memory button M, the average value of
all the stored measured values in the user memory will be
displayed. If you press the memory button M again, the
average value of the morning measurements for the last 7
days will be displayed. (morning: 5.00 – 9.00, display
If you press the memory button M again, the average
value of the evening measurements for the last 7 days will
be displayed. (evening: 18.00 – 20.00, display
continue to press the memory button M, the most recent
individual measured values are displayed in turn with the
date and time.
• You can delete the memory by pressing and holding the
memory button M for 3 seconds.
• To switch off, press the START/STOP button
• If you forget to switch off the unit, it will switch off auto-
matically after 1 minute.

8. Error message/trouble shooting

In the event of errors, the error message
display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely (
appears on the display)
• the pump pressure is higher than 300 mmHg (
appears on the display)
).
• pumping up takes longer than 160 seconds (
on the display)
• there is a system or device error (
). If you
appears on the display)
• the batteries are almost empty
In such cases, repeat the measurement. Ensure that you do
not move or speak.
If necessary, reinsert or replace the batteries.
.
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical speci-
fications", the technical alarm will appear on the display
indicating either " " or "Lo". In such cases, you should seek
_ appears on the
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.