EN-20
Put the calibration mass on the pan, and then press
the Softkey corresponding to the icon
confirmation. To change to a different calibration
point, input the value desired, and then place the
corresponding weight on the pan for calibration.
The message 'Completed' will be displayed on the
screen.
Exit Linearity Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon

3.3.4 GEO Adjustment

Set the GEO factor that corresponds to your location.
GEO codes are numbered 0-31.
Long press the button
Menu. Select the menu item Calibration by
pressing the Softkey corresponding to the icon
.
Scroll to GEO Adjustment using the Softkey
corresponding to the icon
Press the Softkey corresponding to the icon
to edit the GEO value. Press the button
and
enter
the
desired
alphanumeric keypad. After editing, press the
Softkey corresponding to the icon
the menu.
Note: See table 9-4 for GEO values.
for
.
.
to enter the Main
.
value
using
the
to exit