8.3 Sterilizing

9. FUNCTION CHECK

3. Please take note of the following terms:
Interval
Prior to every measurement
with direct skin contact
After every measurement
with direct skin contact
If required
The sterilisation of this device is not permitted.
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• Function check of all the controls shown in the
section entitled "Overview".
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 94.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 94, you may not use the device.
– Have the device repaired by seca service or
by an authorized service partner.
– Please see the section entitled "Servicing" on
page 96.