6.3 Sterilizing
7. FUNCTION CHECK
3. Disinfect the device by moistening a soft cloth in
disinfectant and wiping the device over with it.
:
Interval
Prior to every measurement
with direct skin contact
After every measurement
with direct skin contact
If required
This device may not be sterilised.
►
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview" on page 52
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 76.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 76, you may not use the device.
►
Have the device repaired by seca service or
by an authorized service partner.
►
Please see the section entitled "Servicing/re-
calibration" on page 77.
English
Component
Weighing platform
Weighing platform
Display
Membrane keyboard
Function check • 75