Upstream HEPA filter replacement
Version for dust harmful for the health
WARNING!
Take care not to raise dust when this operation is
carried out. Wear a P3 mask and other protective
clothing plus protective gloves (DPI) suited to the
hazardous nature of the dust collected, refer to the
laws in force.
WARNING!
Do not use the filter again after having removed it from
the vacuum cleaner.
HEPA filter replacement
Figure 14
1.
-
2.
Filter operating knob
3.
Cover
4.
Lever
5.
Safety bolt
6.
Stop nut (absolute filter)
7.
Absolute filter holder disc
8.
Primary filter
9.
Clamp fixing screw
10. Primary filter fixing clamp
11. Absolute filter
12. Vacuum hose
13. Vacuum hose clamp
■
Stop the vacuum cleaner.
■
Loosen the clamp (13) with a screwdriver and remove
the vacuum hose (12).
■
Unscrew the knob (2).
■
Unlock the safety bolt (5).
■
Use the lever (4) to remove the cover (3).
■
Loosen the screw (9) of the primary filter (8) fixing clamp
(10) using a screwdriver.
■
Remove the filter holder disc (7) with the absolute filter
installed and unscrew the nut (6).
■
Remove the absolute filter (11).
■
Cover the absolute filter (11) with a plastic bag, close
the plastic bag hermetically and dispose of the filter in
accordance with the laws in force.
■
Insert a new filter (11) with the same filtering
characteristics as the removed one.
■
Lock the absolute filter with the nut (6).
■
Tighten the clamp (10) screw (9) fixing the primary filter
(8) to the absolute filter holder disc (7).
■
Insert the cover (3) again.
■
Close the cover hermetically by means of the lever (4) and
lock the safety bolt (5) again.
■
Fit the vacuum hose (12) back in place and tighten the
calmp (13).
02/2013
Motor cooling fan inspection and cleaning
These operations can only be carried out by trained
and qualified personnel.
Periodically clean the motor cooling fan to prevent the motor
from overheating, especially if the vacuum cleaner is used in
a dusty place (1, Fig. 4).
Seal inspection
Hoses check
Make sure that connecting hoses are in a good condition and
correctly fixed.
If the hoses are damaged, broken or badly connected to the
unions, they must be replaced.
When sticky materials are treated, check for possible clogging
along the hose (3, Fig. 15), in the inlet and on the baffle plate
inside the filtering chamber.
Scrape inlet (3, Fig. 15) from the outside and remove the
deposited waste as indicated in figure 15.
Filtering chamber seal check
If the gasket (1, Fig. 16) between the container and the
filtering chamber (3) fails to guarantee tightness:
■
Loosen the four screws (2) that lock the filtering chamber
(3) against the vacuum cleaner structure.
■
Allow the filtering chamber (3) to lower down and tighten the
screws (2) once it has reached the tightness position.
If an optimal seal cannot be obtained and the gasket (1) is
torn or cracked, etc. it must be replaced.
Separator cleaning and replacement
(optional)
If there is only a dust deposit on the separator (3, Fig. 17)
allow the dust to drop through the central hole.
The separator (3, Fig. 17) should first be disassembled in
order to be perfectly cleaned:
■
Use the lever (1) to remove the cover (2) together with
the primary filter.
■
Unscrew the two screws (4) and remove it from the
container.
Replace the part if it is excessively worn.
Assemble the components in the reverse order of
disassembly.