10.1 Display's fail to light up

• Check if the mains cable is connected to mains
supply.
• Check if the mains supply is available (power
supply) for instance with a lamp.
• If the lamp does not light up, check the mains
fuses.
• if the lamp lights up, check the fuses of the
unit.

10.2. Error code Er XXX* on the display

The unit has discovered a fault using its "self-test".
Switch the unit back on. If the code re-appears, con-
tact your supplier. The unit is probably defective.
* Er XXX: a number is mentioned on position X

11.1. Microwave therapy

Frequency
HF power, continuous
HF power, pulsed
Operating modes
Clock
26

10. FAULT CONDITIONS

10.3 Power cannot be set

• HF-cable is not connected.
• No radiator is connected.
• No treatment time is set.
10.4 During treatment the HF-power
switches off and a sound signal is
audible ( 3 beeps of 1 sec.)
• The HF-cable is mistakenly disconnected from
radiator or apparatus.
• The HF-cable is damaged and must be replaced.
• Bad contact occurred in the connection of the
HF-cable to the apparatus and/or radiator.
: 2450 MHz ± 50 MHz
: 0 - 250 Watt at 50 Ohm
: 0 - 250 Watt with a fixed peak power of 1500 Watt
+/- 30%
: continuous and pulsed
: 0 - 30 minutes