Fault Conditions; Maintenance By The User - Enraf Nonius Curapuls 970 Instrucciones De Manejo

Ocultar thumbs Ver también para Curapuls 970:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

After switching on the Curapuls 970, the
intensity display and indicator lamp for
tuning does not light.
Check that there is power to the wall socket.
Indicator lamp for tuning does not go out,
but intensity display lights.
Check electrodes and cables, and their connections.
Intensity display does not light, but indicator
lamp for tuning lights.
Check that the automatic cut-out is in position '
22

FAULT CONDITIONS

'.

MAINTENANCE BY THE USER

General
Before cleaning and other maintenance of the
Curapuls 970, the unit must first be switched off, and
disconnected from the mains supply.
Opening of the equipment and attempts at repair by
unauthorized persons can adversely affect the safety
and are therefore not permitted.
Apparatus
The apparatus may be cleaned with a damp cloth
using a non-abrasive, liquid household cleanser. Also
the electrode cables, electrode arms, plate electrodes
and induction electrodes may be cleaned in a similar
way.
Flexible rubber electrodes
Clean the flexible rubber electrodes, preferably daily,
with normal lukewarm water. If necessary, a house-
hold soap may be used. To avoid corrosion of the
cable junction, do not completely submerge the elec-
trodes. Avoid contact of electrodes with grease or
oil.
Before using the flexible rubber electrodes after clean-
ing, make sure that both surfaces are fully dry.
The electrodes may be disinfected with a cloth damp-
ened with 70% alcohol. Be aware that the black of
the rubber pad electrodes may stain.
Electrode cables
Check all cables and electrodes regularly for dam-
age (e.g. minute cracks) and bad contacts. Please
realize that any poor contact might cause a local
overheating effect. Pay special attention to possible
cracks in the insulation of the cables. Always replace
damaged cables or electrodes immediately. We
advise keeping an extra set of frequently used ac-
cessories in stock.
End life span apparatus and accessories
Your Curapuls 970 and accessories contains materi-
als which can be recycled and/or are noxious for the
environment. At the end of the duration of life, spe-
cialised concerns can take apart these articles and
sort out the noxious materials and materials for recy-
cling. By doing so you contribute to a better environ-
ment.
Please ensure that you are well informed of the
local rules and regulations regarding to the re-
moval of apparatus and accessories.