Gima TOBI CLINIC Manual De Istrucciones página 10

Ocultar thumbs Ver también para TOBI CLINIC:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator TOBI CLINIC is intended for use in the electromagnetic environment specified below.
The customers or the user of the surgical aspirator TOBI CLINIC should assure that it's used in such an
Emissions Test
Irradiated / Conducted emissions
CISPR11
Irradiated / Conducted emissions
CISPR11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator TOBI CLINIC is intended for use in the electromagnetic environment specified below.
The customers or the user of the surgical aspirator TOBI CLINIC should assure that it's used in such an
Immunity Test
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
Electrical fast transient / burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
Magnetic field
IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Note U
is the value of the power supply voltage
T
environment.
Compliance
Group 1
Class [B]
Class [A]
Complies
environment.
Compliance
± 6kV on contact
± 8kV in air
± 2kV power supply
± 1kV differential mode
5%U
for 0.5 cycle
T
40%U
for 05 cycle
T
70%U
for 25 cycle
T
<5%U
T
for 5 sec
3A/m
3Vrms 150kHz to 80MHz
(for appliances that aren't life -
supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't life -
equipment)
10
Electromagnetic environment -
guidance
The surgical aspirator
TOBI CLINIC only used RF
energy only for its internal functioning.
Therefore its RF emissions are very low
and are not cause interference in
proximity of any Electronic appliances.
The surgical aspirator
TOBI CLINIC can be used in all
environments, including domestic and
those connected directly to the public
mains distribution that supplies power
to environments used for domestic
scopes.
Electromagnetic environments -
guidance
Floors should be wood, conceret or
ceramic tile. If floors are coverei with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital If the user of the surgical
aspirator TOBI CLINIC
request that the appliance operates
continuosly, the use of a continuity unit
is recommended.
The power frequency magnetic field
should be measured in the intended
installation location to assure that it's
sufficiently low.
-
-

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tobi hospi plusTobi hospital

Tabla de contenido