Figure 2 Installation dimensions

Plumb the sample and drain lines

After the panel is attached to a wall, connect the sample and drain lines to the fittings on the front of
the panel.
1. Connect the sample line to the inlet 1/4-in. OD tube fitting.
2. Connect a drain line to the drain 1/4-in. OD tube fitting. Keep the drain line as short as possible to
prevent back-pressure.
Do not install the drain line back into the main sample stream.

Connect to the controller

The panel connects to an external controller for operation. Connect the cable from the controller to
the quick-connect fitting on the junction box.

Connect the protective earth ground

Connect the ground wire on the flow chamber to a protective earth ground to keep electrical noise to
a minimum. The controller and the panel must use the same source for the protective earth ground.