Operating Instructions SATA AB 1
EN
■ Fill the tank with demineralised water for normal operation and insert
(see chapter 12.1 / 12.2).

14. Care and storage

Attention!
Damage from wrong cleaning agents
The use of aggressive cleaning agents can damage the SATA AB1.
→ Do not use aggressive cleaning agents.
→ Use a neutral cleaning solution with a pH of 6 - 8.
→ Do not use acids, caustic solutions, bases, paint strippers, unsuitable
regenerates or other aggressive cleaning agents.
Careful handling together with regular care of the product is necessary to
warrant the functioning of the SATA AB1.
Clean the SATA AB1 at regular intervals and check for leaks.

15. Malfunctions

Malfunction
Air escapes from the
pressure regulator
valve [4-2] at the
bleeder hole.
Yellow signal lamp
[6-1] "Start / Overtemp.
ack." lights up.

16. Customer service and spare parts

Accessories, spare parts and technical support may be obtained from
your SATA dealer.
54
Cause
Membrane [8-5] de-
fective
Pressure control valve
rod [8-12] defective
Overheating cut-out
of the heating sleeve
has triggered.
Corrective action
Clean and grease
pressure regulator
valve and replace
membrane (see chap-
ter 13.3).
Clean and grease
pressure regulator
valve and replace pres-
sure control valve rod
(see chapter 13.3).
Let heating sleeve cool
down and check that
it is fitted correctly to
the tank. Acknowledge
fault at signal button
"Start / Overtemp. [6-1]
ack.".