Measurement (meas) / Adjustment (cal)
Printing and transferring results of a series of samples manually
- Press and hold the
pears in the display.
- Press the
or
Execute? (
- Press the ok/meas. key to confirm.
- Enter the desired series of samples (from ... to ...) using the arrow keys.
To transfer the series of samples:
- Press the ok/meas. key.
The sample series (from ... to ...) is confirmed. The
results of the selected series of samples are transferred.
To print or transfer all results manually
- Press and hold the
pears in the display.
- Press the ok/meas. key.
The
symbol in the display flashes and all results are transferred.
After a successful transfer, the user is asked whether he/she would like to delete
the transferred results:
Leaving transferred results in tact
- Confirm
Memory All Clear
button.
Deleting transferred results
- Press the
or
Execute? (
- Press the ok/meas. key to confirm.
All results are deleted.
6.8
Deleting results
It is not possible to delete individual results with the Densito 30P.
Deleting all results
- Press and hold the
appears in the display.
- Press the
or
Execute? (
- Press the ok/meas. key to confirm.
All results are deleted.
7
Adjustment (cal)
Before adjusting
- Carry out a test measurement before adjusting the measuring cell, see Chap. 4.3.
Test measurement result <0.9995 or >1.0005
- Check if the measuring cell is dirty, clean if necessary and repeat the test
measurement.
Test measurement result once again <0.9995 or >1.0005
- Readjust instrument.
16
key until
key.
) appears.
Range
key until
Memory All Clear
key.
) appears.
Yes
key until
key.
) appears.
Yes
Memory out
Memory out
Execute? (
Execute? (
No
Memory All Clear
METTLER TOLEDO Densito 30P
, Execute? (
symbol flashes and the
, Execute? (
).
No
) by pressing the ok/meas.
Execute? (
) ap-
All
) ap-
All
)
No