Test Measurement; Adjustment - Mettler Toledo Densito 30P Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4.3

Test measurement

- Push drain button (3) downward completely.
- Ensure that the sample tube is located in the accompanying density standard
or in distilled water.
- Press the fill button (12) slowly to fill the measuring cell.
- Ensure that no air bubbles are contained within the measuring cell.
- Press the ok/meas. key.
The instrument automatically executes a measurement and the result appears in
reverse display.
The result for water must be 1.000 in the selected measurement unit SG (t/t).
If the deviation is < 0.0005
Adjustment is ok, the instrument is ready for measurement.
If the deviation is > 0.0005
Adjust instrument, see Chapter 4.4.
4.4

Adjustment

Initializing adjustment
- Press and hold the
The instrument adjusts automatically (duration: approx. 1 minute). After adjust-
ment is completed, the measured deviation from the theoretical value and
Execute? (
No
- Press the
or
- Press the ok/meas. key to confirm.
The adjustment is confirmed.
METTLER TOLEDO Densito 30P
3
12
key until
CALIB (Water)
) appears.
key. Execute? (
appears on the display.
) appears.
Yes