Calculation of the temperature-compensation coefficient α
- Measure density of the sample:
• at a temperature (T
• at a temperature (T
- Calculate α according to the formula:
density at T
= density at T
2
or
α = ( density at T
density at T
- Enter value a x 1000 into the instrument.
Note
The temperatures T
chapter 5.3).
Example
(measured) density at 26 °C (T
(measured) density at 15 °C (T
α = ( 0.7937 – 1) : (26 – 15)
0.7844
α = (1.011856 – 1) : 11 = 0.011856 : 11 = 0.001078
α
1000 = 1.078; enter this value into the instrument.
x
5.3

Temperature unit (Temp. Unit)

Data in °C or °F (selectable).
5.4

Measure Mode

Configuration of identification, stability mode and method of data storage.
Sample Name
Stability
Auto
Manu
Mode
Labo
Field
Custom User-defined setting.
METTLER TOLEDO Densito 30P
) above the normal measuring temperature,
1
) below the normal measuring temperature.
2
(1 + α
1
– 1 ) : ( T
– T
2
1
1
have to be entered in the selected unit (°C or °F, see
and T
1
2
) 0.7844 g/cm
1
) 0.7937 g/cm
2
Sample identification.
A letter (a...z or space) can be set for the identification of
samples.
Stability control.
The result is accepted automatically when the display is stabile.
Result accepted by pressing the ok/meas. key.
Method of data storage.
By pressing the ok/meas. key the result is saved and tranferred
(printer, PC).
By pressing the ok/meas. key the result is saved.
Memory in
Memory out
( T
– T
))
1
2
)
2
3
3
Save results.
Auto
Save result automatically.
Manu
Result saved by pressing the
ok/meas. key.
Transfer result to PC or printer.
Auto
Transfer result automatically.
Manu
Transfer result by pressing the
key.
Menu (menu)
11