Metrohm 756 KF Instrucciones Para El Uso página 90

Coulometer
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4.2 Remote control commands
82
:
Report configuration. If a report line is switched off, the corresponding line will not
be outputted in the reports.
With "Run" on "OFF", only the run number will not be outputted, date (and time) are
available.
4.2.2.53. Config.ComVar.C30
with up to .C39, etc.
Values of the common variables from C30 up to C39. Insert the common variables
directly or describe the determination results directly from the method, see 4.2.2.24.
4.2.2.54. SmplData.Status
On/off switching of silo memory. When the silo memory is switched on, the sample
data are fetched from the lowest valid silo line.
4.2.2.55. SmplData.OFFSilo.Id1
SmplData.OFFSilo.Id2
SmplData.OFFSilo.Id3
SmplData.OFFSilo.ValSmpl
SmplData.OFFSilo.UnitSmpl
SmplData.OFFSilo.Limits
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calculation
constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be entered.
.Limits:
Limits of sample size of current method.
4.2.2.56. SmplData.ONSilo.Counter.MaxLines
SmplData.ONSilo.Counter.FirstLine
SmplData.ONSilo.Counter.LastLine
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
4.2.2.57. SmplData.ONSilo.EditLine.1.Method
SmplData.ONSilo.EditLine.1.Id1
SmplData.ONSilo.EditLine.1.Id2
SmplData.ONSilo.EditLine.1.Id3
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl
SmplData.ONSilo.EditLine.1.C24
SmplData.ONSilo.EditLine.1.C25
SmplData.ONSilo.EditLine.1.Mark
etc., up to .255
Contents of a silo line.
0... ±999 999
ON, OFF
up to 12 ASCII characters
up to 12 ASCII characters
up to 12 ASCII characters
6-digits, sign and decimal point
up to 5 ASCII characters
read only
read only
read only
read only
up to 8 ASCII characters
up to 12 ASCII characters
up to 12 ASCII characters
up to 12 ASCII characters
up to 5 ASCII characters
read only
read only
read only
756/831 KF Coulometer, Instructions for Use