Invacare Perfecto 2 Serie Manual Del Operador página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 97
SECTION 5—TYPICAL PRODUCT PARAMETERS
Oxygen Output
Concentration
Levels:
NOTE: Stated 
concentration levels 
achieved after initial 
warm‐up period 
(approximately 30 
minutes)
Maximum Outlet
Pressure:
Flow Range:
Low Flow Alarm:
Average Power
Consumption:
Pressure Relief
Mechanism
Operational at:
Change in maximum
recommended flow
when back pressure
of 7kPa is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Perfecto
™Series
2
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
0.5 to 5 L/min (maximum)
For flowrates less than 1 L/min, we recommend the use of the
Invacare Pediatric Flowmeter Accessory (IRCPF16AW)
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories Connected)
Perfecto
AW - 300 W @ 5 L/min, 280 W @ 3 L/min
2
Perfecto
VAW - 320 W@ 5 L/min
2
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown
High temperature compressor shutdown
High Pressure Alarm w/compressor shutdown
Low Pressure Alarm w/compressor shutdown
Battery Free Power Loss Alarm. SensO
Model) Low Flow Alarm.
38.1 cm ± 1 cm (15 in ± 3/8 in)
58.4 cm ± 1 cm (23 in ± 3/8 in)
30.5 cm ± 1 cm (12 in ± 3/8 in)
Perfecto
AW Models - 20.5 kg ± 1 kg (45 lbs ± 2 lbs)
2
Perfecto
VAW Models - 18.1 kg ± 1 kg (40 lbs ± 2 lbs)
2
Perfecto
AW Models - 22.7 kg ± 1 kg (50 lbs ± 2 lbs)
2
Perfecto
VAW Models - 20.4 kg ± 1 kg (45 lbs ± 2 lbs)
2
14
>
Oxygen System (SensO
2
Part No 1163146
2

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Perfecto 2 irc5po2awPerfecto 2 irc5po2vawPerfecto 2 irc5paw

Tabla de contenido