Avoid storage in condensing ambient humidity.
►
Avoid contact with abrasive substances (e.g. sand, dust).
Do not expose the product to temperatures below -10 °C (14 °F) or above +50 °C (122 °F)
►
(e.g. sauna, excessive sunlight, drying on a radiator).
NOTICE
Contact with saltwater or water containing chlorine/soap, fresh water or abrasive sub
stances (e.g. sand)
Damage and premature product wear
►
Following any contact with the substances identified above, promptly clean the product in
accordance with the section "Cleaning".
5 Handling
INFORMATION
►
The daily duration of use and period of application are generally determined by the physi
cian.
►
The initial fitting and application of the product must be carried out by qualified personnel.
►
Instruct the patient in the handling and care of the product.
►
Instruct the patient to see a physician immediately if any exceptional changes are noted (e.g.
worsening of the complaint).
5.1 Selecting the Size
5.2 Adaptation
NOTICE
Improper changes
Damage to the product due to overloading of the material and improper fit of the product due to
breakage of load-bearing components
►
Do not make any improper changes to the product.
The Dyneva is assembled on delivery and must be adapted to the requirements of the patient.
14
>
Required tools: 743S1=60 diameter or
743E10 measuring tool
1) Measure the width of the torso 4 cm below
the shoulder blades (dimension A).
2) Measure the width of the torso at the level
of the iliac crest (dimension B).
3) Determine the size of the tube frames using
the measured dimensions and the size
chart (see Page 3).
4) INFORMATION: Determine the size of
the brace (XS/S/M/L/XL/XXL) on the
basis of the width of the torso below
the shoulder blades.
If necessary, order the lower tube frame
one size smaller or bigger and then
replace
it
29R414=XS/S/M/L/XL/XXL).
(reference
number