Fault-Cause-Remedy

Fault
Cause
Ball sticks after
Static charge
flow has been
terminated
Soiling inside
the flow tube
Cracks and/or
Incorrect
clouding of the
disinfectant or
pressure dome
exposure to
sunlight
Leakage
Missing or
between
worn sealing
flowmeter and
ring at the gas
connected
outlet
accessories

Maintenance

This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
WARNING
Risk of infection
Users and service personnel can become
infected with pathogenic germs.
Disinfect and clean device or device parts
before any maintenance measures and also
before returning the medical device for repair.
14
Definition of maintenance concepts
Concept
Remedy
Disconnect
Maintenance
flowmeter and
set down
briefly
Replace flow
tube, call
Inspection
DrägerService
Replace the
pressure
Preventive
dome. Call
maintenance
DrägerService
Call
Repair
DrägerService
Inspection
Perform inspections at regular intervals and observe
the following specifications.
.
Checks
Inspection
Performing inspection
1
Check completeness and readability of labels
2
Visual inspection of
connectors
gas type code conformity
the sealing ring at the gas outlet
Definition
All measures (inspection,
preventive maintenance, repair)
intended to maintain and restore
the functional condition of a
medical device
Measures intended to determine
and assess the actual state of a
medical device
Recurrent specified measures
intended to maintain the functional
condition of a medical device
Measures intended to restore the
functional condition of a medical
device after a device malfunction
Interval
Personnel
responsible
Every
Service personnel
5 years
the CS connection hose, if available
the pressure dome and the respective seal-
ing ring for cracks and contamination
3
Functional check
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.

Cleaning and disinfection

WARNING
Risk of infection
Reusable products must be reprocessed. Oth-
erwise there is an increased risk of infection
and the functioning of the products may be im-
paired.
Comply with the hospital's hygiene regula-
tions.
Use validated procedures for reprocessing.
Reprocess reusable products before each
use.
Observe the manufacturer's instructions on
cleaning agents and disinfectants.
WARNING
Risk due to faulty products
Reusable products may show signs of wear,
such as cracks, deformations, discolorations,
or peeling.
Check products for signs of wear and replace if
necessary.
Instructions for use Flowmeter